4.5 Article

Calibration methods for VSCs measured on AS-TD-GC-SCD

期刊

出版社

SPRINGER
DOI: 10.1007/s10661-021-09690-y

关键词

Volatile sulfur compounds (VSCs); Sulfur compound standards; Sulfur chemiluminescence detector (SCD); Calibration method

向作者/读者索取更多资源

This study compared three different sample preparation and sampling methods for the calibration of volatile sulfur compounds (VSCs), using both gas and liquid standards. The results showed that the dynamic dilution and loop sampling method achieved the best calibration results.
Volatile sulfur compounds (VSCs) are key odorous compounds from emissions of various odour sources because they are odorous and generally have very low odour threshold values. Identification and quantification of them through air server-thermal desorber-gas chromatography-sulfur chemiluminescence detector (AS-TD-GC-SCD) become more and more popular, although VSCs can be determined by other detectors. To find a valid, practical and quick calibration method is also an important step in their analytical processes. This study compared three different sample preparation and unity sampling methods using both gas standards (with 10 VSCs balanced in pure nitrogen gas) and liquid standards of 7 VSCs. For liquid standard sample preparation, two solvents (methanol and n-pentane) were tested and their calibration results were compared. The study revealed that the three calibration methods with both manual and dynamic dilution of VSC standard gases can achieve satisfactory calibration results with nice linear regression and correlation coefficient (r(2)). The dynamic dilution and loop sampling method is recommended because of its better reliability and time-saving processing. For calibration of VSCs with liquid standards, preparing the samples using dissolved VSCs in n-pentane and analysing them using the loop sampling method achieved best calibration results. For dimethyl trisulfide (DMTS), its calibration cannot obtain as good results as other sulfur compounds even using the best performance calibration method.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据