4.6 Article

Monodispersed nanoplatelets of samarium oxides for biosensing applications in biological fluids

期刊

ELECTROCHIMICA ACTA
卷 402, 期 -, 页码 -

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.electacta.2021.139532

关键词

Nanostructure; Samarium oxide; Electrochemistry; Biosensor; Glucose

资金

  1. Romanian Ministry of Research and Innovation through Operational Programme Competitiveness [NANOBIOSURF-SMIS103528, PN19-030101]

向作者/读者索取更多资源

A nanostructured samarium oxide electrode was prepared by electrodeposition on a gold electrode surface, showing quasi-monodispersed nanoplatelets and coexistence of Sm2+ and Sm3+ species. The electrode displayed improved charge transfer properties and was successfully used for H2O2 detection and glucose biosensing applications with high sensitivity and selectivity.
A nanostructured samarium oxide electrode was constructed by electrodeposition onto the surface of a gold electrode on SiO2/Si wafer. The samarium oxides electrode's surface morphology was investigated by scanning electron microcopy showing a quasi-monodispersed nanoplatelets like structure. X-rays diffraction analysis demonstrated a mixture of monoclinic and hexagonal phases while the X-rays photoelectron spectroscopy indicated the co-existence of both Sm2+ and Sm3+ species in a 1:3 proportion. Cyclic voltammetry and electrochemical impedance spectroscopy were used to investigate the charge transfer processes at the surface of the samarium oxide electrode in the absence and in the presence of redox probes. A roughness factor of 2.5 was determined from the samarium oxide electrode while the charge transfer constant was almost double when compared to the planar gold electrode. Then, the samarium oxide electrode was used for the H2O2 detection by fixed potential amperometry at -0.20 V (vs. Ag/AgCl) with a linear region between 0.01 and 1.00 mM, a sensitivity of 153 mu A cm(-2) mM(-1) and a LoD = 2.70 mu M. Glucose oxidase was used as a model enzyme in order to test the capacity of the samarium oxides electrode for biosensing. The enzyme was immobilized by physical adsorption and the optimum conditions for glucose analysis investigated. The biosensor showed a linear range for glucose detection between 0.10 and 1.20 mM with a sensitivity of 8.40 mu A cm(-2) mM(-1) and a LoD = 8.00 mu M. Selectivity was tested toward common interfering species, and the results revealed the lack of biosensor response. The glucose biosensor on samarium oxides was tested for glucose detection in serum samples with a recovery factor of 90%, and the result validated with a commercial glucometer. (C) 2021 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据