期刊
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS II-EXPRESS BRIEFS
卷 63, 期 12, 页码 1151-1155出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TCSII.2016.2613564
关键词
Data compression; electrocardiogram (ECG)-on-chip; low-power design; QRS detection; wireless sensors
资金
- Singapore National Research Foundation [NRF-CRP8-2011-01]
- Canada Natural Sciences and Engineering Research Council Discovery Grant
This brief presents the design of a low-power 3-lead electrocardiogram (ECG)-on-chip with integrated real-time QRS detection and lossless data compression for wearable wireless ECG sensors. Data compression and QRS detection can reduce the sensor power by up to 2-5 times. A joint QRS detection and lossless data compression circuit allows computational resources to be shared among multiple functions, thus lowering the overall system power. The proposed technique achieves an average compression ratio of 2.15 times on standard test data. The QRS detector achieves a sensitivity (Se) of 99.58% and positive productivity (+P) of 99.57% @ 256 Hz when tested with the MIT/BIH database. Implemented in 0.35 mu m process, the circuit consumes 0.96 mu W @ 2.4 V with a core area of 1.56 mm(2) for two-channel ECG compression and QRS detection. Small size and ultralow-power consumption makes the chip suitable for usage in wearable/ambulatory ECG sensors.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。