Assessment of clinical measures of total and regional body composition from a commercial 3-dimensional optical body scanner

期刊

CLINICAL NUTRITION
卷 41, 期 1, 页码 211-218

出版社

CHURCHILL LIVINGSTONE
DOI: 10.1016/j.clnu.2021.11.031

关键词

3-Dimensional optical; Digital anthropometry; DXA; Fat free mass; Body fat percent; Validation

资金

  1. Shape Up! Adults grant [NIH R01 DK109008]

向作者/读者索取更多资源

This study assessed the accuracy of a commercial 3-dimensional optical imaging system in estimating body composition and anthropometry. The results showed that the measurements provided by this system were accurate and precise compared to criterion methods. This method offers a rapid and automated way to assess body composition regardless of body size.
Background: The accurate assessment of total body and regional body circumferences, volumes, and compositions are critical to monitor physical activity and dietary interventions, as well as accurate disease classifications including obesity, metabolic syndrome, sarcopenia, and lymphedema. We assessed body composition and anthropometry estimates provided by a commercial 3-dimensional optical (3DO) imaging system compared to criterion measures. Methods: Participants of the Shape Up! Adults study were recruited for similar sized stratifications by sex, age (18-40, 40-60, >60 years), BMI (under, normal, overweight, obese), and across five ethnicities (non-Hispanic [NH] Black, NH White, Hispanic, Asian, Native Hawaiian/Pacific Islander). All participants received manual anthropometry assessments, duplicate whole-body 3DO (Styku S100), and dual-energy X-ray absorptiometry (DXA) scans. 3DO estimates provided by the manufacturer for anthropometry and body composition were compared to the criterion measures using concordance correlation coefficient (CCC) and Bland-Altman analysis. Test-retest precision was assessed by root mean square error (RMSE) and coefficient of variation. Results: A total of 188 (102 female) participants were included. The overall fat free mass (FFM) as measured by DXA (54.1 +/- 15.2 kg) and 3DO (55.3 +/- 15.0 kg) showed a small mean difference of 1.2 +/- 3.4 kg (95% limits of agreement-7.0 to +5.6) and the CCC was 0.97 (95% CI: 0.96-0.98). The CCC for FM was 0.95 (95% CI: 0.94-0.97) and the mean difference of 1.3 +/- 3.4 kg (95% CI:-5.5 to +8.1) reflected the difference in FFM measures. 3DO anthropometry and body composition measurements showed high test-retest precision for whole body volume (1.1 L), fat mass (0.41 kg), percent fat (0.60%), arm and leg volumes, (0.11 and 0.21 L, respectively), and waist and hip circumferences (all <0.60 cm). No group differences were observed when stratified by body mass index, sex, or race/ethnicity. Conclusions: The anthropometric and body composition estimates provided by the 3DO scanner are precise and accurate to criterion methods if offsets are considered. This method offers a rapid, broadly available, and automated method of body composition assessment regardless of body size. Further studies are recommended to examine the relationship between measurements obtained by 3DO scans and metabolic health in healthy and clinical populations. Published by Elsevier Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据