4.4 Article

Measurement of the 145Sm half-life

期刊

APPLIED RADIATION AND ISOTOPES
卷 178, 期 -, 页码 -

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2021.109978

关键词

Half-life; Uncertainty; Sm-145; Nuclear data; Brachytherapy

向作者/读者索取更多资源

The half-life of Sm-145 was measured using the reference source method with a HPGe detector, and the result of 345 (16) days is consistent with scarce literature values.
The half-life of Sm-145 has been measured by means of the reference source method with a HPGe detector. The long-lived radionuclide Ti-44 was mixed into the source for reference. The time-dependency of the Sm-145/Ti-44 activity ratio was followed by assessing the count-rate ratio of their characteristic gamma-ray emissions at 61.2 keV (Sm-145) and 67.9/78.3 keV (Ti-44) in spectra recorded over periods of typically one day. In total, 220 measurements were performed over a period of 384 days or about one half-life period. The experiment and ensuing uncertainty budget are discussed in detail. Different error propagation is applied for random uncertainties, autocorrelated structures in the fit residuals, and potential systematic errors. The result for the Sm-145 half-life, 345 (16) d, is compatible with the scarce literature values, however the experimental details of the old measurements were barely documented.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据