4.8 Article

Novel Self-Calibrating Amperometric and Ratiometric Electrochemical Nanotip Microsensor for pH Measurement in Rat Brain

期刊

ANALYTICAL CHEMISTRY
卷 93, 期 41, 页码 13815-13822

出版社

AMER CHEMICAL SOC
DOI: 10.1021/acs.analchem.1c02436

关键词

-

资金

  1. National Science Foundation of China [22074089, 22104078]
  2. Chinese Postdoctoral Science Foundation [2020M672258]
  3. Central Thousand Talents Plan [ZYQR201810151]
  4. Key Scientific Research Projects of Colleges and Universities in Henan Province [21A150043]
  5. East China Normal University

向作者/读者索取更多资源

A novel voltammetric pH sensor was developed by measuring the ratio of current signals, providing accurate in situ pH detection within the range of 4.5 to 8.2. The sensor demonstrated excellent selectivity, high repeatability, and stability, successfully detecting pH fluctuations in rat cerebral ischemia and whole blood. This research not only offers a valuable tool for rat brain pH detection, but also presents a new strategy for designing nanosensors for intracellular or subcellular pH measurements.
Brain pH has been proven to be a key factor in maintaining normal brain function. The relationship between local pH fluctuation and brain disease has not been extensively studied due to lack of the accurate in situ analysis technology. Herein, we have for the first time proposed a voltammetric pH sensor by measuring the ratio of current signals instead of the previously reported potential based on the Nernst equation. Single-walled carbon nanotubes (CNT) were first self-assembled on the electrode surface of a carbon-fiber nanotip electrode (CFNE). Then, poly-o-phenylenediamine (PoPD) molecules were deposited as pH-responsive molecules through in situ electrochemical polymerization. The compact CFNE/CNT/PoPD exhibited a good redox process with the on-off-on ratiometric electrochemical response to pH ranging from 4.5 to 8.2, providing self-correction for in situ pH detection. Thus, the proposed sensor enabled the accurate measurement of pH with excellent selectivity even in the presence of proteins or electroactive species. In addition, the sensor showed high repeatability, reproducibility, and reversibility in measuring pH and even demonstrated good stability when it was exposed to air for 5 months. Finally, we successfully detected the fluctuation of pH in rat brains with cerebral ischemia and rat whole blood. Overall, this research not only provides a good tool for the detection of rat brain pH but also provides a new strategy for further designing nanosensors for intracellular or subcellular pH.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据