4.7 Article

Pseudo isotope dilution (PID) as an approach for correcting barium-related spectral interferences on the measurement of europium by inductively coupled plasma mass spectrometry (ICP-MS)

期刊

ANALYTICA CHIMICA ACTA
卷 1180, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.aca.2021.338854

关键词

Pseudo isotope dilution; Spectral interference; Barium; Europium; ICP-MS

资金

  1. Japan Society for the Promotion of Science (JSPS) [19K12300]
  2. Grants-in-Aid for Scientific Research [19K12300] Funding Source: KAKEN

向作者/读者索取更多资源

A practical method called pseudo isotope dilution (PID) was proposed to correct barium-related spectral interferences on the measurement of europium. Results confirmed its validity by determining europium in standard solutions with different concentrations of barium.
Pseudo isotope dilution (PID) was suggested as a simple and practical approach for the correction of barium-related spectral interferences on the measurement of europium by inductively coupled plasma mass spectrometry (ICP-MS). The interferents, i.e. oxides and hydroxides of barium, were regarded as pseudo europium isotopes. An equation for calculating the concentration of europium was deduced and the net contribution of europium isotopes can be obtained by measuring the signal intensity ratios of I-151/I-153 in a sample, a europium standard, and a barium standard, respectively. The validity of the approach was confirmed by the determination of europium in standard solutions with different concentrations of barium. Determination of europium in river water and seawater certified reference materials was carried out, whose results were in coincidence with reference values. (C) 2021 Published by Elsevier B.V.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据