4.5 Article

Monitoring of Lactate in Interstitial Fluid, Saliva and Sweat by Electrochemical Biosensor: The Uncertainties of Biological Interpretation

期刊

CHEMOSENSORS
卷 9, 期 8, 页码 -

出版社

MDPI
DOI: 10.3390/chemosensors9080195

关键词

lactate biosensor; Pediococcus sp lactate oxidase; subcutaneous lactate monitoring; salivary lactate; sweat lactate; exercise

资金

  1. UK EPSRC [EP/H009744/1]
  2. EPSRC [EP/H009744/1] Funding Source: UKRI

向作者/读者索取更多资源

Lactate electrochemical biosensors were successfully fabricated using Pediococcus sp lactate oxidase, enabling detection of lactate in subcutaneous tissue, saliva, and sweat. The study revealed significant differences between blood lactate and salivary lactate, suggesting limitations in using lactate sensors for extravascular monitoring.
Lactate electrochemical biosensors were fabricated using Pediococcus sp lactate oxidase (E.C. 1.1.3.2), an external polyurethane membrane laminate diffusion barrier and an internal ionomeric polymer barrier (sulphonated polyether ether sulphone polyether sulphone, SPEES PES). In a needle embodiment, a Pt wire working electrode was retained within stainless steel tubing serving as pseudoreference. The construct gave linearity to at least 25 mM lactate with 0.17 nA/mM lactate sensitivity. A low permeability inner membrane was also unexpectedly able to increase linearity. Responses were oxygen dependent at pO(2) < 70 mmHg, irrespective of the inclusion of an external diffusion barrier membrane. Subcutaneous tissue was monitored in Sprague Dawley rats, and saliva and sweat during exercise in human subjects. The tissue sensors registered no response to intravenous Na lactate, indicating a blood-tissue lactate barrier. Salivary lactate allowed tracking of blood lactate during exercise, but lactate levels were substantially lower than those in blood (0-3.5 mM vs. 1.6-12.1 mM), with variable degrees of lactate partitioning from blood, evident both between subjects and at different exercise time points. Sweat lactate during exercise measured up to 23 mM but showed highly inconsistent change as exercise progressed. We conclude that neither tissue interstitial fluid nor sweat are usable as surrogates for blood lactate, and that major reappraisal of lactate sensor use is indicated for any extravascular monitoring strategy for lactate.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据