4.7 Article

Macular Pigment Reflectometry: Developing Clinical Protocols, Comparison with Heterochromatic Flicker Photometry and Individual Carotenoid Levels

期刊

NUTRIENTS
卷 13, 期 8, 页码 -

出版社

MDPI
DOI: 10.3390/nu13082553

关键词

macular pigment optical density; MPOD; lutein; zeaxanthin; carotenoids; heterochromatic flicker photometry; age related macular degeneration; biomarker

向作者/读者索取更多资源

The study aimed to analyze measurement protocols using Macular Pigment Reflectometry (MPR), compare MPOD measurements from MPR and heterochromatic flicker photometry (MPS II), and assess lutein and zeaxanthin optical density in the central macula. Results showed high correlation and agreement between MPR-MPOD and MPS II-MPOD values, with the 10-30 interval providing the most accurate measurements. Dilation improved repeatability of lutein and zeaxanthin values, and MPR was able to generate unique optical density data not available from other clinical devices.
The study was designed to: (1) Analyze and create protocols of obtaining measurements using the Macular Pigment Reflectometry (MPR). (2) To assess the agreement of MPOD measurements obtained using the heterochromatic flicker photometry (MPS II) and MPR. (3) To obtain the lutein and zeaxanthin optical density obtained using the MPR in the central one-degree of the macula. The measurements were performed using the MPR and heterochromatic flicker photometry. The MPR measurements were performed twice without pupillary dilation and twice following pupillary dilation. The MPR measurements were performed for a 40-s period and the spectrometer signal was parsed at different time points: 10-20, 10-30, 10-40, 20-30, 20-40, and 30-40 s. The MPR analyzes the high-resolution spectrometer signal and calculates MPOD, lutein optical density and zeaxanthin optical density automatically. The MPR-MPOD data was compared with MPPS II-MPOD results. The MPR-MPOD values are highly correlated and in good agreement with the MPS II-MPOD. Of the various parsing of the data, the data 10-30 interval was the best at obtaining the MPOD, lutein, and zeaxanthin values (8-12% coefficient of repeatability). The lutein to zeaxanthin ratio in the central one-degree of the macula was 1:2.40. Dilation was not needed to obtain the MPOD values but provided better repeatability of lutein and zeaxanthin optical density. MPR generates MPOD measurements that is in good agreement with MPS II. The device can produce lutein and zeaxanthin optical density which is not available from other clinical devices.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据