Investigating the position of the separation capillary and emitter tube tips in a nanoflow sheath-liquid CE-ESI-MS interface to decouple the ESI potential

期刊

TALANTA
卷 228, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2021.122212

关键词

Capillary electrophoresis; Electrospray ionization; ESI-MS interface; ESI potential decoupling; Mass spectrometry; Spray potential

资金

  1. Czech Science Foundation [20-03899S]
  2. Czech Academy of Sciences [61388963]

向作者/读者索取更多资源

In CE-ESI-MS interfaces, protruding or aligning the capillary tip with the emitter tip affects the decoupling of the ESI potential from the separation potential. Only when the capillary tip is retracted 20 μm inside the emitter can complete decoupling of the spray and separation potentials be achieved, providing better sensitivity for the analysis.
Robust decoupling of the ESI potential from the separation potential in CE-ESI-MS interfaces is very important for the high performance of the CE-ESI-MS devices and their applications for highly sensitive analyses of ionogenic compounds. In this study, we utilize a nanoflow sheath-liquid CE-ESI-MS interface composed of a quartz emitter and a separation fused silica capillary treated by etching, which are threaded to cross coupling for sheath liquid and electrode connection. Specifically, we have tested the ability of the interface to decouple the ESI potential from the separation potential at different positions of the separation capillary and ESI emitter tube tips. The interface with the separation capillary tip protruding the emitter tip by 20 mu m did not provide sufficient robustness. The real ESI potential (delivered as 2.0 kV from the independent high voltage power supply HV2) ranged from 2.1 kV to 4.5 kV depending on the applied separation voltage (12.0-20.0 kV, provided by the power supply HV1) and electric conductivity of the background electrolyte (BGE) used. The interface robustness was partially improved when the capillary tip was aligned with the emitter tip. However, the complete decoupling of the spray and separation potentials was achieved only when the capillary tip was retracted 20 mu m inside the emitter. In this arrangement, the ESI potential was stable and independent of both the separation potential (voltage) and the BGE conductivity. Moreover, this setting provided better sensitivity for the CE-ESI-MS analysis of selected drugs and benzylpyridinium cations than the setup with the capillary tip aligned with or protruding the emitter tip.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据