4.6 Article

New Calibration System for Low-Cost Suspended Particulate Matter Sensors with Controlled Air Speed, Temperature and Humidity

期刊

SENSORS
卷 21, 期 17, 页码 -

出版社

MDPI
DOI: 10.3390/s21175845

关键词

calibration; low-cost sensors; air quality; particulate matter

资金

  1. Jagiellonian University
  2. Faculty of Physics, Astronomy and Applied Computer Science of the Jagiellonian University in Krakow

向作者/读者索取更多资源

The paper introduced a calibration system for low-cost PM sensors developed and used at Jagiellonian University in Krakow, Poland, which is capable of generating stable and repeatable concentrations of suspended PM.
This paper presents a calibration system for low-cost suspended particulate matter (PM) sensors, consisting of reference instruments, enclosed space in a metal pipe (volume 0.145 m(3)), a duct fan, a controller and automated control software. The described system is capable of generating stable and repeatable concentrations of suspended PM in the air duct. In this paper, as the final result, we presented the process and effects of calibration of two low-cost air pollution stations-university measuring stations (UMS)-developed and used in the scientific project known as Storm&DustNet, implemented at the Jagiellonian University in Krakow (Poland), for the concentration range of PM from a few up to 240 mu g center dot m(-3). Finally, we postulate that a device of this type should be available for every system composed of a large number of low-cost PM sensors.