4.6 Article

DMAS Beamforming with Complementary Subset Transmit for Ultrasound Coherence-Based Power Doppler Detection in Multi-Angle Plane-Wave Imaging

期刊

SENSORS
卷 21, 期 14, 页码 -

出版社

MDPI
DOI: 10.3390/s21144856

关键词

delay-and-sum (DAS); delay-multiply-and-sum (DMAS); signal coherence; power doppler detection; plane-wave (PW) imaging; complementary subset transmit (CST); coherent plane-wave compounding (CPWC)

资金

  1. Ministry of Science and Technology of Taiwan [108-2221-E-011-072-MY3, 108-2221-E-011-071-MY3]

向作者/读者索取更多资源

In this study, DMAS beamforming technique is used to enhance Doppler SNR in ultrafast acquisition. Simulation results show that DMAS can significantly improve Doppler SNR compared to conventional DAS beamforming. Further improvement can be achieved by combining DMAS with CST technique and sufficient ensemble averaging.
Conventional ultrasonic coherent plane-wave (PW) compounding corresponds to Delay-and-Sum (DAS) beamforming of low-resolution images from distinct PW transmit angles. Nonetheless, the trade-off between the level of clutter artifacts and the number of PW transmit angle may compromise the image quality in ultrafast acquisition. Delay-Multiply-and-Sum (DMAS) beamforming in the dimension of PW transmit angle is capable of suppressing clutter interference and is readily compatible with the conventional method. In DMAS, a tunable p value is used to modulate the signal coherence estimated from the low-resolution images to produce the final high-resolution output and does not require huge memory allocation to record all the received channel data in multi-angle PW imaging. In this study, DMAS beamforming is used to construct a novel coherence-based power Doppler detection together with the complementary subset transmit (CST) technique to further reduce the noise level. For p = 2.0 as an example, simulation results indicate that the DMAS beamforming alone can improve the Doppler SNR by 8.2 dB compared to DAS counterpart. Another 6-dB increase in Doppler SNR can be further obtained when the CST technique is combined with DMAS beamforming with sufficient ensemble averaging. The CST technique can also be performed with DAS beamforming, though the improvement in Doppler SNR and CNR is relatively minor. Experimental results also agree with the simulations. Nonetheless, since the DMAS beamforming involves multiplicative operation, clutter filtering in the ensemble direction has to be performed on the low-resolution images before DMAS to remove the stationary tissue without coupling from the flow signal.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据