3D sodium (23Na) magnetic resonance fingerprinting for time-efficient relaxometric mapping

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 86, 期 5, 页码 2412-2425

出版社

WILEY
DOI: 10.1002/mrm.28873

关键词

7 Tesla; Cramer Rao lower bound; magnetic resonance fingerprinting; relaxometry; sodium; X-nuclei

资金

  1. Projekt DEAL

向作者/读者索取更多资源

A framework for 3D sodium MR fingerprinting was developed, allowing simultaneous quantification of T1, T2l*, T2s*, and T2*. The method used tailored flip angle patterns and time-efficient data acquisition. Experimental results in both phantoms and human brains demonstrated the feasibility of this approach to sodium relaxometric mapping within approximately 30 minutes.
Purpose: To develop a framework for 3D sodium (Na-23) MR fingerprinting (MRF), based on irreducible spherical tensor operators with tailored flip angle (FA) pattern and time-efficient data acquisition for simultaneous quantification of T-1, T-2l*, T-2s*, and T-2* in addition to Delta B-0. Methods: Na-23-MRF was implemented in a 3D sequence and irreducible spherical tensor operators were exploited in the simulations. Furthermore, the Cramer Rao lower bound was used to optimize the flip angle pattern. A combination of single and double echo readouts was implemented to increase the readout efficiency. A study was conducted to compare results in a multicompartment phantom acquired with MRF and reference methods. Finally, the relaxation times in the human brain were measured in four healthy volunteers. Results: Phantom experiments revealed a mean difference of 1.0% between relaxation times acquired with MRF and results determined with the reference methods. Simultaneous quantification of the longitudinal and transverse relaxation times in the human brain was possible within 32 min using 3D Na-23-MRF with a nominal resolution of (5 mm)(3). In vivo measurements in four volunteers yielded average relaxation times of: T-1,T-brain = (35.0 +/- 3.2) ms, T-2l,T-brain* = (29.3 +/- 3.8) ms and T-2s,T-brain* = (5.5 +/- 1.3) ms in brain tissue, whereas T-1,T-CSF = (61.9 +/- 2.8) ms and T-2,T-CSF* = (46.3 +/- 4.5) ms was found in cerebrospinal fluid. Conclusion: The feasibility of in vivo 3D relaxometric sodium mapping within roughly 1/2 h is demonstrated using MRF in the human brain, moving sodium relaxometric mapping toward clinically relevant measurement times.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据