The study found that varus alignment of the tibial component up to 7 degrees did not negatively impact implant survival, patient satisfaction, and function of a neutrally aligned primary TKA, with a minimum follow-up of 10 years.
Purpose The purpose of this retrospective study was to evaluate the effect of varus alignment of the tibial component on the outcomes with a minimum follow-up of 10 years. The hypothesis was that varus alignment of the tibial component might not affect the outcomes and survival of a neutrally aligned primary TKA. Methods A matched case-control study was designed between 66 patients with varus alignment of the tibial component and 66 with neutral alignment with a minimum follow-up of 10 years. Functional outcome was assessed with the knee surgery scores (KSS) and reduced Western Ontario and MacMaster Universities questionnaire (WOMAC). Patient satisfaction was evaluated by a 0-10 visual analog scale. Radiological evaluation was performed at early postoperative and at final follow-up. Results The mean follow-up was 11.9 (SD 2.6) years for both groups. The mean postoperative proximal tibial angle in the varus group was 85.0 degrees (SD 0.9) and 88.8 degrees (SD 0.9) in control group. At the final follow-up, there were no significant differences in KSS, WOMAC, range of motion or patient satisfaction. There were no differences in the coronal anatomical alignment of the TKA between groups. Revision of TKA was performed in four knees in the varus group, and one in control group, due to aseptic loosening of the tibial component in all cases. TKA survival at 10 years was not significantly different between groups. Conclusion The alignment of the tibial component up to 7 degrees varus did not negatively affect implant survival, patient satisfaction, and function of a well-aligned TKA, with a minimum postoperative follow-up of 10 years.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据