4.6 Article

Accurately calibrated frequency domain diffuse optical spectroscopy compared against chemical analysis of porcine adipose tissue

期刊

JOURNAL OF BIOPHOTONICS
卷 14, 期 12, 页码 -

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jbio.202100169

关键词

calibration; chemical analysis; diffuse optical spectroscopy; porcine adipose; tissue lipid; tissue water

资金

  1. Ministry of Science and ICT and MOE [NRF-2018K1A4A3A02060572, NRF-2019R1I1A3A01062141, NRF-2020H1D3A1A04080958]

向作者/读者索取更多资源

Frequency domain diffuse optical spectroscopy (fdDOS) is a noninvasive technique used to estimate tissue composition and hemodynamics, and can be calibrated using liquid and solid phantoms for comparison with direct chemical analysis (CA). Results showed average errors and standard deviations of 1.9 +/- 0.2% and -0.9 +/- 0.2% for water content, and -9.3 +/- 1.3% and 0.8 +/- 1.3% for fat content measured by fdDOS compared to CA.
Frequency domain diffuse optical spectroscopy (fdDOS) is a noninvasive technique to estimate tissue composition and hemodynamics. While fdDOS has been established as a valuable modality for clinical research, comparison of fdDOS with direct chemical analysis (CA) methods has yet to be reported. To compare the two approaches, we propose a procedure to confirm accurate calibration by use of liquid emulsion and solid silicone phantoms. Tissue fat (FAT) and water (H2O) content of two ex vivo porcine tissue samples were optically measured by fdDOS and compared to CA values. We show an average H2O error (fdDOS minus CA) and SD of 1.9 +/- 0.2% and -0.9 +/- 0.2% for the two samples. For FAT, we report a mean error of -9.3 +/- 1.3% and 0.8 +/- 1.3%. We also measured various body sites of a healthy human subject using fdDOS with results suggesting that accurate calibration may improve device sensitivity.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据