4.4 Article

How to choose the suitable FNS specification in young patients with femoral neck fracture: A finite element analysis

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.injury.2021.05.043

关键词

Femoral neck fracture; Femoral neck system; Cannulated screw; Finite element analysis; FEA

资金

  1. National Natural Science Foundation of China (CN) [82004387]
  2. Natural Science Foundation of Guangdong Province [2018A030313694]

向作者/读者索取更多资源

The optimal approach for the treatment of femoral neck fractures remains debated. A new internal fixation femoral neck system (FNS) was developed and tested for biomechanical outcomes in unstable fractures. Results indicate that single-hole or double-hole locking plated FNS can be used for fractures with angles less than 60 degrees, while a double-hole locking plated FNS is recommended for angles greater than 70 degrees.
Background: Consensus regarding the optimal approach for the treatment of femoral neck fractures remains lacking. A new internal fixation femoral neck system (FNS) was developed and used in clinical practice. We aimed to investigate the biomechanical outcomes of different types of FNS in the treatment of unstable femoral neck fractures. Method: In this study, we constructed three different types of unstable femoral neck fractures of Pauwels classification with angles of 50 degrees, 60 degrees, and 70 degrees. We set up four test groups, namely, the one-hole plated FNS group, two-hole plated FNS group, inverted cannulated screw group and triangle cannulated screw group. Under 2100 N axial loads, displacements and the von Mises stress of the femur and internal fixation components were measured for each fracture group. Results: When the Pauwels angle was 50 degrees or 60 degrees, the one-hole locking plated FNS was as superior as the two-hole plated FNS in terms of femur and internal fixation displacement, and the inverted cannulated screw had slightly better stability than the triangular cannulated screw. However, when the angle increases to 70 degrees, the two-hole locking plate has the minimum displacement, followed by the triangular cannulated screw and inverted cannulated screw, which is the worst displacement for the single-hole locking plate. Regardless of the angle, the two sets of FNS have higher internal fixation stress than the two sets of cannulated screws, which is approximately 1.6-3.0 times that of the cannulated screw group. Conclusion: From the perspective of biomechanics, we suggest that when the angle of the fracture line is less than 60 degrees, both single-hole locking plated or double-hole locking plated FNS can be used to treat unstable femoral neck fractures. However, when the angle of the fracture line is greater than 70 degrees, we recommend using a double-hole locking plated FNS. This result needs further verification in further clinical studies. (C) 2021 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据