Design and Construction of a PET-Compatible Double-Tuned 1H/31P MR Head Coil

期刊

IEEE TRANSACTIONS ON MEDICAL IMAGING
卷 40, 期 8, 页码 2015-2022

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TMI.2021.3070626

关键词

Radio frequency; Protons; Capacitors; Imaging; Absorption; Magnetic resonance imaging; Head; H-1; P-3(1); double-tuned RF coil; low count-rate loss; simultaneous PET-MR

向作者/读者索取更多资源

The study demonstrated a PET-transparent MR coil system for a double-tuned H-1/P-31 head coil at 3T, achieving minimal count rate loss and impact on PET image quality through proper component arrangement and design optimization. The MR-related performance was comparable or slightly better than that of a commercial double-tuned coil.
Simultaneous MR-PET is an increasingly popular multimodal imaging technique that is able to combine metabolic information obtained from PET with anatomical/functional information from MRI. One of the key technological challenges of the technique is the integration of a PET-transparent MR coil system, a solution to which is demonstrated here for a double-tuned H-1/P-31 head coil at 3 T. Two single-resonant birdcage coils tuned to the H-1 and P-31 resonances were arranged in an interleaved fashion and electrically decoupled with the use of trap circuits. All high 511 keV quanta absorbing components were arranged outside the PET field-of-view in order to minimize count rate reduction. The materials inside the PET field-of-view were carefully evaluated and chosen for minimum impact on the PET image quality. As far as possible, the coil case was geometrically optimized to avoid sharp transitions in attenuation, which may potentially result in streaking artefacts during PET image reconstruction. The coil caused a count rate loss of just above 5% when inserted into the PET detector ring. Except for the anterior region, which was designed to maintain free openings for increased patient comfort, an almost uniform distribution of 511 keV attenuation was maintained around the circumference of the coil. MR-related performance for both nuclei was similar or slightly better than that of a commercial double-tuned coil, despite the MR-PET coil having a close-fitting RF screen to shield the PET and MR electronics from possible electromagnetic interferences.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据