Calibration Technique for THz Time-Domain Spectrometers Enabling Vectorial Scattering Parameter Measurements

期刊

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/LMWC.2021.3062679

关键词

Calibration; Scattering parameters; Wireless communication; Extraterrestrial measurements; Time-domain analysis; Semiconductor device measurement; Measurement by laser beam; Calibration; millimeter-wave measurements

资金

  1. Menlo Systems GmbH

向作者/读者索取更多资源

The most proficient instrument for rapid measurements of broadband terahertz spectra is the time-domain spectrometer, driven by a strong need for fast and precise instrumentation in application-oriented research. This study proposes an adapted network analysis calibration scheme for magnitude- and phase-correct wideband scattering parameter measurements, validated through a comparison of simulation and measurement results.
To date, the most proficient instrument type for rapid measurements of broadband terahertz (THz) spectra is the time-domain spectrometer, which acquires the RF signal either transmitted through or reflected by a sample without following a standardized method for calibration. Simultaneously, application-oriented research drives a strong need for fast and precise instrumentation capable of capturing well-established characteristics, such as scattering parameters. In this letter, a robust THz instrument capable of synchronously measuring transmission and reflection from 150 GHz to 3 THz is implemented and evaluated. The main sources of error in the system are characterized, and an adapted network analysis calibration scheme for magnitude- and phase-correct wideband scattering parameter measurements is proposed. Finally, the method is validated in a comparison of simulation and measurement of a known sample.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据