4.2 Article

Capillary Electrophoresis-Indirect Laser-Induced Fluorescence Detection of Neomycin in Fish

期刊

CHROMATOGRAPHIA
卷 84, 期 9, 页码 861-868

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s10337-021-04075-2

关键词

Capillary electrophoresis; Indirect laser-induced fluorescence detection; Neomycin; Fish

资金

  1. National Natural Science Foundation of China (NSFC) [21005021]
  2. Natural Science Foundation of Guangdong Province [2016A030313740, 2021A1515011410]
  3. Guangdong Provincial Science and Technology Project [2016B030303002]

向作者/读者索取更多资源

A capillary electrophoresis-indirect laser-induced fluorescence detection method was established for neomycin detection in fish. With optimal conditions, the limit of detection of neomycin was 5.0 ng/g, the recovery of fish samples is between 95.2 and 99.7%, and the relative standard deviation is between 2.7 and 3.6%. The study also investigated the effects of background buffer, pH, and separation voltage on the separation and analysis process.
A capillary electrophoresis-indirect laser-induced fluorescence detection method was established for neomycin detection in fish. With rhodamine 6G as the background fluorescent substance, neomycin with weak ultraviolet absorption can be detected effectively. The types and concentrations of background buffer, pH, and separation voltage which affected the separation and analysis were all studied. With the optimal conditions, the limit of detection of neomycin was 5.0 ng/g, the recovery of fish samples is between 95.2 and 99.7%, and the relative standard deviation is between 2.7 and 3.6%.