A systematic review of the sensitivity and specificity of lateral flow devices in the detection of SARS-CoV-2

期刊

BMC INFECTIOUS DISEASES
卷 21, 期 1, 页码 -

出版社

BMC
DOI: 10.1186/s12879-021-06528-3

关键词

Coronavirus; COVID-19; SARS-CoV-2; Lateral flow device; Lateral flow test; Viral antigen detection; Rapid antigen detection; Reverse transcriptase polymerase chain reaction; Mass testing; Population testing

向作者/读者索取更多资源

This systematic review found that the performance of lateral flow devices is heterogeneous and dependent on the manufacturer. Some devices have high specificity but a great range of sensitivities were shown (38.32-99.19%). Test performance does not appear to be dependent on the operator.
Background Lateral flow devices (LFDs) are viral antigen tests for the detection of SARS-CoV-2 that produce a rapid result, are inexpensive and easy to operate. They have been advocated for use by the World Health Organisation to help control outbreaks and break the chain of transmission of COVID-19 infections. There are now several studies assessing their accuracy but as yet no systematic review. Our aims were to assess the sensitivity and specificity of LFDs in a systematic review and summarise the sensitivity and specificity of these tests. Methods A targeted search of Pubmed and Medxriv, using PRISMA principles, was conducted identifying clinical studies assessing the sensitivity and specificity of LFDs as their primary outcome compared to reverse transcriptase polymerase chain reaction (RT-PCR) for the detection of SARS-CoV-2. Based on extracted data sensitivity and specificity was calculated for each study. Data was pooled based on manufacturer of LFD and split based on operator (self-swab or by trained professional) and sensitivity and specificity data were calculated. Results Twenty-four papers were identified involving over 26,000 test results. Sensitivity from individual studies ranged from 37.7% (95% CI 30.6-45.5) to 99.2% (95% CI 95.5-99.9) and specificity from 92.4% (95% CI 87.5-95.5) to 100.0% (95% CI 99.7-100.0). Operation of the test by a trained professional or by the test subject with self-swabbing produced comparable results. Conclusions This systematic review identified that the performance of lateral flow devices is heterogeneous and dependent on the manufacturer. Some perform with high specificity but a great range of sensitivities were shown (38.32-99.19%). Test performance does not appear dependent on the operator. Potentially, LFDs could support the scaling up of mass testing to aid track and trace methodology and break the chain of transmission of COVID-19 with the additional benefit of providing individuals with the results in a much shorter time frame.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据