4.8 Article

Synovial Fluid pH Sensor for Early Detection of Prosthetic Hip Infections

期刊

ADVANCED FUNCTIONAL MATERIALS
卷 31, 期 37, 页码 -

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/adfm.202104124

关键词

hip infections; hydrogels; pH; plain radiographies; synovial fluid

资金

  1. NIH [R01 AR070305-01, P30 GM131959]
  2. Wallace R. Roy Professorship

向作者/读者索取更多资源

An implantable sensor has been developed to measure synovial fluid pH for noninvasive early detection and monitoring of hip infections using standard-of-care plain radiography. The sensor shows a linear response and reversibility in the physiologically relevant pH range.
An implantable sensor developed to measure synovial fluid pH for noninvasive early detection and monitoring of hip infections using standard-of-care plain radiography is described. The sensor is made of a pH responsive polyacrylic acid-based hydrogel, which expands at high pH and contracts at low pH. A radiodense tantalum bead and a tungsten wire are embedded in the two ends of the hydrogel to monitor the change in length of the hydrogel sensor in response to pH via plain radiography. The effective acid dissociation constant (pK(a)) of the hydrogel-based pH sensor is 5.6 with a sensitivity of 3 mm/pH unit between pH 4 and 8. The sensor shows a linear response and reversibility in the physiologically relevant pH range of pH 6.5 and 7.5 in both buffer and bovine synovial fluid solutions with a 30-minute time constant. The sensor is attached to an explanted prosthetic hip, and the pH response is determined from the X-ray images by measuring the length between the tantalum bead and the radiopaque wire. Therefore, the developed sensor will enable noninvasive detection and studying of implant hip infection using plain radiography.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据