Phthalate Exposures in the Neonatal Intensive Care Unit

期刊

TOXICS
卷 9, 期 5, 页码 -

出版社

MDPI
DOI: 10.3390/toxics9050090

关键词

hypertension; prematurity; phthalates; di-2-ethylhexyl phthalate (DEHP); toxicity

资金

  1. Friends of Doernbecher Foundation
  2. Juliet A. Hillman Foundation

向作者/读者索取更多资源

The study shows that in a neonatal unit, infants receive 98% of DEHP exposure from respiratory devices, with bubble CPAP accounting for 95% of the total exposure. By avoiding the use of IV tubing containing DEHP and modifying respiratory equipment appropriately, DEHP exposure can be significantly reduced.
Background: Di-2-ethylhexyl phthalate (DEHP), a phthalate compound found in medical devices, may cause toxic effects in premature infants. In this study, the objective is to quantify DEHP exposures from various intravenous and respiratory therapy devices, and to use these values to predict typical exposure for an infant in a neonatal unit. Methods: Common IV products used on infants are directed through various types of IV tubing (IVT) and analyzed for DEHP content. DEHP exposure for infants receiving respiratory therapy was determined indirectly through analysis of urine DEHP metabolites. By deriving these values for DEHP we calculated the daily exposure to DEHP from common IV fluids (IVF) and respiratory devices during hospitalization in a neonatal unit. Results: IVF labeled DEHP-positive showed very high concentrations of DEHP, but when passed through IVT, substantial amounts were adsorbed. DEHP was undetectable with all DEHP-negative IVF tests, except when passed through DEHP-positive IVT. The DEHP leached from most respiratory devices was relatively modest, except that detected from bubble CPAP. In 14 very low birthweight infants, the mean DEHP exposure was 182,369 mcg/kg over 81.2 days of the initial hospitalization. Ninety-eight percent of the exposure was from respiratory devices, with bubble CPAP accounting for 95% of the total DEHP exposure in these infants. Conclusions: The DEHP exposure in our neonatal unit can be reduced markedly by avoiding or modifying bubble CPAP equipment and avoiding IV tubing containing DEHP.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据