期刊
TALANTA
卷 233, 期 -, 页码 -出版社
ELSEVIER
DOI: 10.1016/j.talanta.2021.122446
关键词
ICP-MS; Fe-Mn oxides; Trace elements; Rare earth elements; High-resolution; Tm spike
资金
- TOTAL
- IFREMER
An optimal analytical procedure specific to ferromanganese (Fe-Mn) oxides was proposed through investigating different modes of data acquisition. The results showed that high-resolution acquisition mode is unnecessary to achieve high quality data for Rare Earth Elements (REE) in Fe-Mn oxides.
In order to propose an optimal analytical procedure specific to ferromanganese (Fe-Mn) oxides, we investigated different modes of data acquisition using inductively coupled plasma mass spectrometry (ICP-MS). The results of trace element and Rare Earth Element (REE) determination in eight Fe-Mn nodules and crusts (FeMn-1, GSMC-1, GSMC-2, GSMC-3, GSPN-2, GSPN-3, NOD-A-1 and NOD-P-1) are presented here. The analytical procedure involves chemical dissolution of the Fe-Mn oxides and addition of a thulium (Tm) spike. The correction of measured values from potential isobaric interferences was investigated using both corrections based on monoelemental solutions, and data acquisition in the high-resolution mode. The obtained results show that the high-resolution acquisition mode is unnecessary to achieve high quality data for REE in Fe-Mn oxides. Using our revised method, we provide a consistent set of precise and accurate values for eight widely used but poorly characterized certified reference materials.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐