4.7 Article

A sensitive spectrophotometric ellipsometry based Aptasensor for the vascular endothelial growth factor detection

期刊

TALANTA
卷 225, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2020.121982

关键词

Aptamers; Biosensors; Breast cancer; Spectroscopic ellipsometry; VEGF

资金

  1. Scientific Research Projects Division of Bilecik Seyh Edebali University (BSEU-BAP, Bilecik, Turkey) [2019-01.BSEU.03-03]

向作者/读者索取更多资源

A sensitive and selective sensor based on aptamer and spectroscopic ellipsometry was developed for early diagnosis of breast cancer. The sensor achieved a detection range of VEGF165 from 1 pM to 1000 pM in buffer, with detection limits of 5.81 pM and 4.29 pM, respectively.
A sensitive and selective, aptamer and spectroscopic ellipsometry based sensor is reported here for the early diagnosis of breast cancer, which is a common type of cancer following lung cancer. It was aimed to develop a single-step and label-free assay for the sensitive and selective detection of VEGF165. To this end, two different aptamers and spectroscopic ellipsometry were used. In the presented study, by determining the appropriate aptamer immobilization conditions, the spectroscopic ellipsometry technique was successfully applied for the detection of VEGF165 at the range of 1 pM-1000 pM in the buffer. Aptasensors have a detection limit of 5.81 pM and 4.29 pM, respectively.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据