Ultrahigh dose-rate (FLASH) x-ray irradiator for pre-clinical laboratory research

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 66, 期 9, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1361-6560/abf2fa

关键词

FLASH radiotherapy; ultrahigh dose rate irradiation; pre-clinical irradiation system; orthovoltage x-rays; rotating anode x-ray sources; kV x-rays

向作者/读者索取更多资源

FLASH irradiation has been proven to reduce normal tissue toxicity significantly while maintaining tumor control probability compared to conventional irradiation. A novel self-shielded x-ray irradiation cabinet system has been proposed to enhance preclinical research capabilities, utilizing two high capacity 150 kVp fluoroscopy x-ray sources in a parallel-opposed arrangement.
FLASH irradiation has been shown to reduce significantly normal tissue toxicity compared to conventional irradiation, while maintaining tumor control probability at similar level. Clinical translation of FLASH irradiation necessitates comprehensive laboratory studies to elucidate biological effects as well as pertinent technological and physical requirements. At present, FLASH research employs complex accelerator technologies of limited accessibilities. Here, we study the feasibility of a novel self-shielded x-ray irradiation cabinet system, as an enabling technology to enhance the preclinical research capabilities. The proposed system employs two commercially available high capacity 150 kVp fluoroscopy x-ray sources with rotating anode technology in a parallel-opposed arrangement. Simulation was performed with the GEANT4 Monte-Carlo platform. Simulated dosimetric properties of the x-ray beam for both FLASH and conventional dose-rate irradiations were characterized. Dose and dose rate from a single kV x-ray fluoroscopy source in solid water phantom were verified with measurements using Gafchromic films. The parallel-opposed x-ray sources can deliver over 50 Gy doses to a 20 mm thick water equivalent medium at ultrahigh dose-rates of 40-240 Gy s(-1). A uniform depth-dose rate (+/- 5%) is achieved over 8-12 mm in the central region of the phantom. Mirrored beams minimize heel effect of the source and achieve reasonable cross-beam uniformity (+/- 3%). Conventional dose-rate irradiation (<= 0.1 Gy s(-1)) can also be achieved by reducing the tube current and increasing the distance between the phantom and tubes. The rotating anode x-ray source can be used to deliver both FLASH and conventional dose-rate irradiations with the field dimensions well suitable for small animal and cell-culture irradiations. For FLASH irradiation using parallel-opposed sources, entrance and exit doses can be higher by 30% than the dose at the phantom center. Beam angling can be employed to minimize the high surface doses. Our proposed system is amendable to self-shielding and enhance research in regular laboratory setting.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据