Characterization of displacement forces and image artifacts in the presence of passive medical implants in low-field (<100 mT) permanent magnet-based MRI systems, and comparisons with clinical MRI systems

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2021.04.003

关键词

Low-field MRI; Image artifacts; Displacement forces; MR safety; Passive implants

资金

  1. Horizon 2020 European Research Grant [FET-OPEN 737180 Histo MRI]
  2. Simon Stevin Meester Prize
  3. Horizon 2020 ERC [NOMA-MRI 670629]

向作者/读者索取更多资源

The study investigated the displacement forces and image artifacts associated with passive medical implants in low-field MRI systems (<100 mT) compared with higher field strengths. Results showed that displacement forces on all implants except the ferromagnetic one were between 1 and 10 mN at 50 mT, and image artifacts were much less at low-field compared to clinical field strengths. SAR simulations demonstrated that long echo train turbo spin echo sequences can be run within SAR limits under certain conditions.
Purpose: To investigate the displacement forces and image artifacts associated with passive medical implants for recently-developed low-field (<100 mT) MRI systems, and to compare these with values from higher field strengths used for clinical diagnosis. Methods: Setups were constructed to measure displacement forces in a permanent magnet-based Halbach array used for in vivo MRI at 50 mT, and results compared with measurements at 7 T. Image artifacts were assessed using turbo (fast) spin echo imaging sequences for four different passive medical implants: a septal occluder, iliac stent, pedicle screw and (ferromagnetic) endoscopic clip. Comparisons were made with artifacts produced at 1.5, 3 and 7 T. Finally, specific absorption rate (SAR) simulations were performed to determine under what operating conditions the limits might be approached at low-field. Results: Displacement forces at 50 mT on all but the ferromagnetic implant were between 1 and 10 mN. Image artifacts at 50 mT were much less than at clinical field strengths for all passive devices, and with the exception of the ferromagnetic clip. SAR simulations show that very long echo train (>128) turbo spin echo sequences can be run with short inter-pulse times (5-10 ms) within SAR limits. Conclusions: This work presents the first evaluation of the effects of passive implants at field strengths less than 100 mT in terms of displacement forces, image artifacts and SAR. The results support previous claims that such systems can be used safely and usefully in challenging enviroments such as the intensive care unit.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据