The performance of human cytomegalovirus digital PCR reference measurement procedure in seven external quality assessment schemes over four years

期刊

METHODS
卷 201, 期 -, 页码 65-73

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.ymeth.2021.03.016

关键词

Digital PCR; Human cytomegalovirus; External quality assessment; Reference measurement procedure

资金

  1. EMRP project - EMRP participating countries within EURAMET [HLT08]
  2. INFECTMET project [HLT08]
  3. Anti-MicroResist project
  4. European Union [HLT08]
  5. Slovenian Research Agency [15HLT07, P4-0165]
  6. Ministry of Education, Science and Sport of republic of Slovenia [1000-13-0105]
  7. UK National Measurement System
  8. Metrology Institute of the Republic of Slovenia (MIRS)
  9. European Regional Development Fund
  10. [3330-14-500133]

向作者/读者索取更多资源

The study evaluated a candidate digital PCR (dPCR)-based reference measurement procedure for quantification of human cytomegalovirus (hCMV) in multiple schemes, demonstrating high reproducibility and trueness of the method. The dPCR has the potential to act as a calibration independent reference measurement procedure for value assignment of hCMV calibration and support qPCR calibration.
A candidate digital PCR (dPCR)-based reference measurement procedure for quantification of human cytomegalovirus (hCMV) was evaluated in 10 viral load comparison schemes (seven external quality assessment (EQA) and three additional training schemes) organized by INSTAND e.V. over four years (between September 2014 and March 2018). Four metrology institutes participated in these schemes using the same extraction method and dPCR measurement procedure for the hCMV specific target sequence of UL54 gene. The calibration independent reference measurement procedure results from the metrology institutes were compared to the results of the clinical diagnostic laboratories applying hCMV qPCR measurement procedures calibrated to reference materials. While the criteria for the acceptable deviation from the target value interval for INSTAND's EQA schemes is from-0.8 log10 to +0.8 log(10), the majority of dPCR results were between-0.2 log10 to +0.2 log(10). Only 4 out of 45 results exceeded this interval with the maximum deviation of-0.542 log(10). In the training schemes containing samples with lower hCMV concentrations, more than half of the results deviated less than +/- 0.2 log(10) from the target value, while more than 95% deviated less than +/- 0.4 log(10) from the target value. Evaluation of intra-and inter-laboratory variation of dPCR results confirmed high reproducibility and trueness of the method. This work demonstrates that dPCR has the potential to act as a calibration independent reference measurement procedure for the value assignment of hCMV calibration and reference materials to support qPCR calibration as well as ultimately for routine hCMV load testing.