Performance Evaluation of the Siemens SARS-CoV-2 Total Antibody and IgG Antibody Test

期刊

LABORATORY MEDICINE
卷 52, 期 6, 页码 E147-E153

出版社

OXFORD UNIV PRESS
DOI: 10.1093/labmed/lmab027

关键词

COVID-19; SARS-CoV-2; serology; antibody kinetics; performance evaluation; immunoassay

向作者/读者索取更多资源

This study evaluated the performance of 2 commercially available SARS-CoV-2 antibody assays and found both tests to have acceptable performance. Although Seroconversion occurred earlier with the COV2T test, both tests showed a sensitivity approaching 100% after 14 days of symptom onset.
Objective: In this study, the performance of 2 commercially available SARS-CoV-2 antibody assays is evaluated. Methods: The Siemens SARS-CoV-2 Total (COV2T) and IgG (COV2G) antibody tests were evaluated on a Siemens Atellica IM1300 analyzer. Imprecision was assessed with the CLSI EP15 protocol using positive controls. Ninety control group specimens were analyzed for specificity, and 175 specimens from 58 patients with polymerase chain reaction-confirmed SARS-CoV-2 were measured for the sensitivity and kinetics of the antibbody response. Results: Within-run and total imprecision were acceptable for both assays. Both tests showed a specificity of 100%. Sensitivity earlier in the disease state was greater for the COV2T assay than for the COV2G assay, but sensitivity >14 days after onset of symptoms approached 100 degrees h for both. For all patients, antibody titers remained above the seroconversion cutoff for all follow-up specimens. Conclusion: This study shows acceptable performance for both the Siemens COV2T and COV2G test, although seroconversion occurs earlier with the COV2T test.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.9
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据