JOURNAL OF MEDICAL VIROLOGY
卷 93, 期 9, 页码 5655-5659出版社
WILEY
DOI: 10.1002/jmv.27094
关键词
antigen; BioSpeedia; COVID-19; Panbio; RT-qPCR; SARS-CoV-2
类别
This study evaluated the performances of two rapid antigen testing devices in detecting SARS-CoV-2 in nasopharyngeal samples, showing relatively high sensitivity, especially with lower cycle threshold values, and 100% specificity for both devices.
The current reliable recommended test for coronavirus disease 2019 (COVID-19) diagnosis is quantitative reverse-transcription polymerase chain reaction (RT-qPCR). Rapid antigen test devices could be useful as they are less expensive, faster without the need of specialized laboratories to perform the test. We report the performances of two rapid immunochromatographic antigen testing devices compared with RT-qPCR for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection in nasopharyngeal samples. We carried out a lateral-flow tests study on 401 nasopharyngeal swab samples from nonduplicated suspected COVID-19 subjects. An equal volume of universal transport medium tubes-containing samples (dilution ratio = 1:15) were added to the manufacturer's extraction buffer solution (dilution ratio = 1:2) and analyzed on BioSpeedia COVID19Speed-Antigen Test and on Abbott Panbio (TM) COVID-19 Ag Rapid Test, devices. Qualitative results were compared to those obtained by the RT-qPCR (Allplex (TM) SARS-CoV-2 Assay Seegene). Based on our data, the overall sensitivity for BioSpeedia and Panbio devices was estimated at 65.5% and 75.0%, respectively. The sensitivity was greater for cycle threshold values less than 25 achieving 90.4 and 96.8 for BioSpeedia and Panbio devices, respectively. A perfect specificity of 100.0% was observed for both devices.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据