4.6 Article

Femoral Neck Notching in Dual Mobility Implants: Is This a Reason for Concern?

期刊

JOURNAL OF ARTHROPLASTY
卷 36, 期 8, 页码 2843-2849

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2021.03.043

关键词

dual mobility; impingement; femoral neck notching; dislocation; revision

向作者/读者索取更多资源

This study identified femoral notching in 3.5% of cases with dual mobility total hip arthroplasty, mainly associated with modular cylindrospheric cobalt-chromium DM implants. The notches were mostly located in the anterior and superior part of the prosthetic femoral neck. These notches did not result in stem fracture or require revision surgery so far.
Background: Dual mobility (DM) total hip arthroplasty (THA) implants have been advocated for patients at risk for impingement due to abnormal spinopelvic mobility. Impingement against cobalt-chromium acetabular bearings, however, can result in notching of titanium femoral stems. This study investigated the incidence of femoral stem notching associated with DM implants and sought to identify risk factors. Methods: A multicenter retrospective study reviewed 256 modular and 32 monoblock DM components with minimum 1-year clinical and radiographic follow-up, including 112 revisions, 4 conversion THAs, and 172 primary THAs. Radiographs were inspected for evidence of femoral notching and to calculate acetabular inclination and anteversion. Revisions and dislocations were recorded. Results: Ten cases of femoral notching were discovered (3.5%), all associated with modular cylindrospheric cobalt-chromium DM implants (P = .049). Notches were first observed radiographically at mean 1.3 years after surgery (range 0.5-2.7 years). Notch location was anterior (20%), superior (60%), or posterior (20%) on the prosthetic femoral neck. Notch depth ranged from 1.7% to 20% of the prosthetic neck diameter. Eight cases with notching had lumbar pathology that can affect spinopelvic mobility. None of these notches resulted in stem fracture, at mean 2.7-year follow-up (range 1-7.6 years). There were no dislocations or revisions in patients with notching. Conclusion: Femoral notching was identified in 3.5% of DM cases, slightly surpassing the dislocation rate in a cohort selected for risk of impingement and instability. Although these cases of notching have not resulted in catastrophic failures thus far, further study of clinical sequelae is warranted. Component position, spinopelvic mobility, and implant design may influence risk. (C) 2021 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据