Photon-Counting Computed Tomography for Coronary Stent Imaging In Vitro Evaluation of 28 Coronary Stents

期刊

INVESTIGATIVE RADIOLOGY
卷 56, 期 10, 页码 653-660

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/RLI.0000000000000787

关键词

photon-counting; detector; computed tomography; coronary artery stents; cardiac

向作者/读者索取更多资源

This study evaluated the in-stent lumen visibility and quantitative image characteristics of different coronary stents using a novel PCD CT system compared to a conventional EID CT scanner. The results showed that the PCD system provided superior in-stent lumen visibility and image characteristics compared to the EID system, demonstrating its potential for improving imaging quality in coronary interventions.
Objectives The aim of this study was to assess in-stent lumen visibility and quantitative image characteristics of different coronary stents using a novel photon-counting detector (PCD) computed tomography (CT) system in comparison to a state-of-the-art energy-integrating detector (EID) CT scanner. Materials and Methods In this in vitro phantom study, 28 different coronary stents ranging from 2.25 to 4.5 mm lumen diameter were expanded into plastic tubes filled with contrast agent. Stent-containing plastic tubes were positioned in a custom-made emulsion-filled phantom, which was inserted into an anthropomorphic phantom simulating a medium-sized patient. Computed tomography scans were acquired parallel to the scanners' z axis using a novel cadmium telluride-based PCD CT system (SOMATOM CountPlus; Siemens Healthcare GmbH, Forchheim Germany), operating in 2 different modes (standard-resolution mode [SR] and ultra-high-resolution [UHR] mode), and a latest generation dual-source EID CT system (SOMATOM Force; Siemens Healthcare GmbH, Forchheim). CTDIvol-matched images were reconstructed with comparable convolution kernels and using the same reconstruction parameters. In-stent lumen visibility (in %), increase in in-stent attenuation (expressed as Delta in-stent CT attenuation), and image noise (in Hounsfield unit) were manually measured. Parts of the image analysis (in-stent lumen visibility) were additionally performed in an automated way. Differences were tested using Wilcoxon signed rank test. Results The best in-stent lumen visibility was achieved with the PCD-UHR mode and the lowest noise levels with the PCD-SR mode. The median in-stent lumen visibility was significantly higher (P < 0.001) with PCD (SR, 66.7%; interquartile range [IQR], 63.3-72.3; UHR, 68.9%; IQR, 64.4-74.4) compared with EID (65.4%; IQR, 62.2-70.4). The Delta in-stent CT attenuation was significantly lower for PCD in both SR (78 HU; IQR, 46-108; P = 0.024) and UHR (85 HU; IQR, 59-113; P = 0.006) compared with EID (108 HU; IQR, 85-126). Image noise was significantly lower (P < 0.001) for PCD-SR (21 HU; IQR, 21-21) compared with EID images (25 HU; IQR, 24-25.0). Conclusions The PCD provides superior in-stent lumen visibility and quantitative image characteristics when compared with conventional EID.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据