Prospective study of 1308 nasopharyngeal swabs from 1033 patients using the LUMIPULSE SARS-CoV-2 antigen test: Comparison with RT-qPCR

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ijid.2021.02.005

关键词

SARS-CoV-2; COVID-19; Antigen; RT-qPCR; LUMIPULSE

资金

  1. Yamanashi Prefecture
  2. Japan Society for the Promotion of Science KAKENHI [JP18K16292]
  3. YASUDA Medical Foundation
  4. Uehara Memorial Foundation
  5. Medical Research Grants from Takeda Science Foundation
  6. [20H03668]

向作者/读者索取更多资源

This prospective study demonstrated that the LUMIPULSE antigen test is highly accurate for SARS-CoV-2 detection, showing high sensitivity and specificity, particularly in diagnosing the disease accurately in the late phase of infection.
Background: Reverse transcription polymerase chain reaction (RT-PCR) is the gold standard for detection of severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2). Previously, the accuracy of the quantitative LUMIPULSE SARS-CoV-2 antigen test was demonstrated using samples collected retrospectively. In this study, the LUMIPULSE antigen test was clinically validated using prospective samples. Methods: In total, 1033 nasopharyngeal swab samples were collected from 1033 individuals, and an additional 275 follow-up samples were collected from 43 patients who subsequently tested positive for coronavirus disease 2019 (COVID-19). All 1308 samples were subjected to quantitative RT-PCR (RT-qPCR) and the antigen test. The antibody response was investigated for patients with discordant results to clarify if seroconversion had occurred. Results: RT-qPCR identified 990 samples as negative and 43 as positive, while the antigen test identified 992 samples as negative, 37 as positive and four as inconclusive. The overall concordance rate was 99.7% (1026/1029). Sensitivity, specificity, positive predictive value and negative predictive value of the antigen test were 92.5% (37/40), 100% (989/989), 100% (37/37) and 99.7% (989/992), respectively, after exclusion of the four inconclusive results. The kappa coefficient was 0.960 (95% confidence interval 0.892-0.960), suggesting excellent agreement between the two tests. Seropositivity in five of seven patients with discordant results suggested that the discrepancy was caused by samples collected during the late phase of infection. Using follow-up samples, correlation was observed between the antigen level and the viral load or cycle threshold value. The concordance rate between these test results tended to be high among samples collected 0-9 days after symptom onset, but this decreased gradually in samples collected thereafter. Conclusions: This prospective study demonstrated that the LUMIPULSE antigen test is a highly accurate diagnostic test for SARS-CoV-2. (C) 2021 The Author(s). Published by Elsevier Ltd on behalf of International Society for Infectious Diseases.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据