Longitudinal agreement of four bioimpedance analyzers for detecting changes in raw bioimpedance during purposeful weight gain with resistance training

期刊

EUROPEAN JOURNAL OF CLINICAL NUTRITION
卷 75, 期 7, 页码 1060-1068

出版社

SPRINGERNATURE
DOI: 10.1038/s41430-020-00811-3

关键词

-

资金

  1. Biospace Co., Ltd, DBA InBody [C14618]

向作者/读者索取更多资源

The study suggests that different bioimpedance analyzers have absolute differences in detecting bioelectrical changes, which should not be used interchangeably, especially when reference values are utilized. Despite absolute differences, analyzers with varying characteristics demonstrated similar abilities to detect changes in R, Xc, and phi over time.
Background Due to inherent errors involved in the transformation of raw bioelectrical variables to body fluids or composition estimates, the sole use of resistance (R), reactance (Xc), and phase angle (phi) has been advocated when quantifying longitudinal changes. The aim of this investigation was to assess the ability of four bioimpedance analyzers to detect raw bioimpedance changes induced by purposeful weight gain with resistance training. Methods Twenty-one resistance trained males completed a 6-week lifestyle intervention with the aim of purposeful weight gain. Bioimpedance analysis was performed before and after the intervention using four different analyzers (MFBIA(InBody): InBody 770; MFBIA(SECA): Seca mBCA 515/514; BIS: ImpediMed SFB7; SFBIA: RJL Quantum V) for the quantification of R, Xc, and phi at the 50-kHz frequency. Repeated measures ANOVA and follow up tests were performed. Results Analysis revealed main effects of time and method for R, Xc, and phi (p <= 0.02), without significant time x method interactions (p >= 0.07). Follow up for time main effects indicated that, on average, R decreased by 4.5-5.8%, Xc decreased by 2.3-4.0%, and phi increased by 1.8-2.6% across time for all analyzers combined. However, varying levels of disagreement in absolute values were observed for each bioelectrical variable. Conclusions The differences in absolute bioelectrical values suggests that analyzers should not be used interchangeably, which holds particular importance when reference values are utilized. Despite absolute differences, analyzers with varying characteristics demonstrated similar abilities to detect changes in R, Xc, and phi over time.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据