4.5 Article

Evaluation of residual contamination on reused healing abutments

期刊

CLINICAL ORAL INVESTIGATIONS
卷 25, 期 10, 页码 5889-5895

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00784-021-03894-9

关键词

Healing abutments; Residual contamination; Sterilization; Reuse; Dental implant

向作者/读者索取更多资源

This study evaluated the residual contamination of healing abutments sterilized by dealers of dental implant companies. Approximately 31% of the healing abutments showed visible residual contamination. A significant relationship was found between implant brands and contamination status, indicating that reusing healing abutments is not suitable due to incomplete removal of contaminants.
Objectives The purpose of this study is to evaluate residual contamination of healing abutments sterilized by dealers of dental implant companies. Material and methods Healing abutments collected from dealers of different implant companies were randomly numbered, and they were examined for visible deformation and contamination on the abutment surface, packaging errors, and presence of a sterilization indicator. The healing abutments were stained with a dye solution for contamination detection and evaluated by dividing into different regions. The presence of residual contamination was examined through photographic images taken from the healing abutments and the data obtained were recorded. Pearson's chi-square analysis was used to evaluate the relation between implant brands and contamination status. Significance was evaluated as p Results Visible residual contamination was detected in approximately 31% of the healing abutments. After the staining procedure, in the visual assessment of the stained surfaces, residual contamination was found in at least one region of all healing abutments except one (n=176, 99.4%). The most intense staining was seen in the screw driver hole (n=172, 97.1%), body (n=113, 63.8%), occlusal (n=107, 60.4%), screw (n=79, 44.6%), and connector (n=72, 40.6%) regions, respectively. A significant relationship was found between the brands and the contamination status (p <0.05). Conclusion The reuse of healing abutments is not suitable, as it does not ensure optimum cleaning and sterilization, and it is not possible to eliminate contaminants completely.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据