Verification of UriSed 3 PRO automated urine microscope in regional laboratory environment

期刊

CLINICA CHIMICA ACTA
卷 515, 期 -, 页码 96-103

出版社

ELSEVIER
DOI: 10.1016/j.cca.2021.01.005

关键词

Automation; Particle counting; Poisson distribution; Urinalysis; Urine sediment; UriSed 3 PRO

向作者/读者索取更多资源

The study validated the UriSed 3 PRO automated microscopes and found high accuracy and consistency in white blood cell, red blood cell, and epithelial cell counts when compared with visual microscopy and urine culture. The instruments also demonstrated high sensitivity in detecting bacteria.
Background and aims: Ten UriSed 3 PRO automated microscopes (77 Elektronika, Hungary) were verified for nine HUSLAB laboratories with 160 000 annual urine samples. Materials and methods: Particle counting of the primary UriSed 3 PRO instrument (77 Elektronika, Hungary) was verified against reference visual microscopy with 463 urine specimens, and against urine culture on chromogenic agar plates with parallel 396 specimens. Nine secondary instruments were compared pairwise with the primary instrument. Results: Relative imprecisions compared to Poisson distribution, R(CV), were estimated to be 1.0 for white blood cell (WBC) and 1.5 for red blood cell (RBC) counts, respectively. Spearman's correlations against visual microscopy were rS = 0.94 for WBC, rS = 0.87 for RBC, and rS = 0.82 for squamous epithelial cell (SEC) counts. Agreement with visual microscopy (Cohen's weighted kappa) was 0.94 for WBC, 0.89 for RBC, 0.88 for SEC, 0.59 for combined casts, and 0.49 for non-squamous epithelial cells (NEC). Bacteria were detected with a sensitivity of 90% and specificity of 39 against culture at 107 CFB/L (104 CFU/mL). Created flagging limits allowed automated reporting for 70-75% of patient results. Conclusions: UriSed 3 PRO instruments were adopted into routine use after acceptance of the verification.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据