4.5 Article

Comparison of mastectomy and breast reconstruction outcomes using low thermal dissection versus traditional electrocautery: a blinded randomized trial

期刊

BREAST CANCER RESEARCH AND TREATMENT
卷 188, 期 1, 页码 101-106

出版社

SPRINGER
DOI: 10.1007/s10549-021-06177-9

关键词

Breast cancer; Mastectomy; Breast reconstruction; Electrocoagulation; Pain; Drainage; Trial

类别

资金

  1. Medtronic

向作者/读者索取更多资源

The study aimed to compare mastectomy flap perfusion using PlasmaBlade (R) versus traditional electrocautery, with 90% of patients showing better perfusion on the PlasmaBlade (R) side. Although the PlasmaBlade (R) group had higher intraoperative skin flap perfusion ratings, both groups had comparable outcomes.
Introduction PlasmaBlade (R) is a thermal dissection device that may allow for improved perfusion of the mastectomy flap by limiting thermal injury. In this study we aim to compare the mastectomy flap perfusion using PlasmaBlade (R) versus traditional electrocautery. Methods Patients undergoing bilateral mastectomy with immediate breast reconstruction were recruited. The right and left breasts of each patient were randomized to dissection with PlasmaBlade (R) or standard electrocautery. Randomization was performed using random sequences on the day of surgery and was blinded to the plastic surgeon. Mastectomy flap perfusion was assessed following completion of the mastectomy using intra-operative fluoroscopy and plastic surgeon review. Surgical site drainage and pain score were measured. Sign tests were employed to assess differences in perfusion and Wilcoxon paired test for the secondary outcomes. Results Twenty patients were enrolled in the study with median age of 40.5 years and median BMI of 26 kg/m(2). In 18 patients (90%), perfusion was assessed to be better on the side of the PlasmaBlade (R) dissection. Median daily drainage over a 7-day period was 51 cc (IQR 35-61) on the PlasmaBlade (R) side and 44 cc (IQR 31-61) on the control side. Median pain score on the PlasmaBlade (R) side was 4.0 (IQR 2.3-5.9) and 4.4 (IQR 2.9-6) on the control side. No skin necrosis was noted in either groups. Conclusion Use of PlasmaBlade (R) appears to be a safe and reliable technique to perform mastectomy and breast reconstruction with equivalent outcomes to traditional electrocautery. Although, mastectomy skin flap perfusion was rated better intra-operatively for the PlasmaBlade (R) group, both cohorts had comparable outcomes. ClinicalTrials.gov Identifier: NCT03711916

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据