4.8 Article

Antifouling Peptide Hydrogel Based Electrochemical Biosensors for Highly Sensitive Detection of Cancer Biomarker HER2 in Human Serum

期刊

ANALYTICAL CHEMISTRY
卷 93, 期 19, 页码 7355-7361

出版社

AMER CHEMICAL SOC
DOI: 10.1021/acs.analchem.1c01350

关键词

-

资金

  1. National Natural Science Foundation of China [21974075]
  2. Science and Technology Benefiting the People Project of Qingdao [20-3-453 nsh]
  3. Taishan Scholar Program of Shandong Province of China [ts20110829]

向作者/读者索取更多资源

The study presents a facile strategy for the electrochemical detection of HER2 using an antifouling sensing interface based on PEDOT and a peptide hydrogel, showing high sensitivity and selectivity for HER2 detection in complex biological samples.
A facile strategy for the electrochemical detection of human epidermal growth factor receptor 2 (HER2), a breast cancer biomarker, was presented via the fabrication of an antifouling sensing interface based on the conducting polymer poly(3,4-ethylenedioxythiophene) (PEDOT) and a biocompatible peptide hydrogel. The peptide hydrogel prepared from a designed short peptide of Phe-Glu-Lys-Phe functionalized with a fluorene methoxycarbonyl group (Fmoc-FEKF) enabled excellent activity preservation for the immobilized biomolecules, and its good hydrophilicity facilitated effective alleviation of nonspecific adsorption or biofouling, while the PEDOT film provided a highly stable and conducting substrate. The developed biosensor was highly sensitive and selective for HER2 detection, with a wide linear response range from 0.1 ng mL(-1) to 1.0 mu g mL(-1) and a low limit of detection of 45 pg mL(-1). Moreover, the peptide hydrogel based biosensor was feasible to use for complex biological samples, and it was capable of detecting HER2 in human serum with clinically acceptable accuracy, manifesting a promising potential for practical application.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据