Theranostic SPECT reconstruction for improved resolution: application to radionuclide therapy dosimetry

期刊

EJNMMI PHYSICS
卷 8, 期 1, 页码 -

出版社

SPRINGER
DOI: 10.1186/s40658-021-00362-x

关键词

Radionuclide therapy; Tumour dosimetry; Theranostics; Partial volume correction

资金

  1. Sydney Vital Translational Cancer Research Centre (Cancer Institute NSW)
  2. EPSRC [EP/T026693/1, EP/M022587/1] Funding Source: UKRI

向作者/读者索取更多资源

The SPECTRE approach to image reconstruction in theranostic setting shows promise in improving accuracy of radioactivity concentrations and treatment assessment by utilizing PET images to guide higher-resolution SPECT reconstructions. Further investigation and optimization of the algorithm parameters are needed before clinical implementation.
BackgroundSPECT-derived dose estimates in tissues of diameter less than 3x system resolution are subject to significant losses due to the limited spatial resolution of the gamma camera. Incorporating resolution modelling (RM) into the SPECT reconstruction has been proposed as a possible solution; however, the images produced are prone to noise amplification and Gibbs artefacts. We propose a novel approach to SPECT reconstruction in a theranostic setting, which we term SPECTRE (single photon emission computed theranostic reconstruction); using a diagnostic PET image, with its superior resolution, to guide the SPECT reconstruction of the therapeutic equivalent. This report demonstrates a proof in principle of this approach.MethodsWe have employed the hybrid kernelised expectation maximisation (HKEM) algorithm implemented in STIR, with the aim of producing SPECT images with PET-equivalent resolution. We demonstrate its application in both a dual Ga-68/Lu-177 IEC phantom study and a clinical example using Cu-64/Cu-67.ResultsSPECTRE is shown to produce images comparable in accuracy and recovery to PET with minimal introduction of artefacts and amplification of noise.ConclusionThe SPECTRE approach to image reconstruction shows improved quantitative accuracy with a reduction in noise amplification. SPECTRE shows great promise as a method of improving SPECT radioactivity concentrations, directly leading to more accurate dosimetry estimates in small structures and target lesions. Further investigation and optimisation of the algorithm parameters is needed before this reconstruction method can be utilised in a clinical setting.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据