4.4 Article

Utility of an Automatic Vision-Based Examination System (AVE-562) for the Detection of Clonorchis sinensis Eggs in Stool

期刊

ANNALS OF LABORATORY MEDICINE
卷 41, 期 2, 页码 221-224

出版社

KOREAN SOC LABORATORY MEDICINE
DOI: 10.3343/alm.2021.41.2.221

关键词

AVE-562 analyzer; Performance; Parasite; Clonorchis sinensis; Eggs

资金

  1. Chonnam National University [2018-3309]
  2. Chonnam National University Hwasun Hospital Institute for Biomedical Science [HCRI 19026]
  3. Basic Science Research Program through the National Research Foundation of Korea (NRF) - Ministry of Education [NRF-2019R1C1C1004605, NRF-2019M3E5D1A02067953]
  4. National Research Foundation of Korea [2019M3E5D1A02067953] Funding Source: Korea Institute of Science & Technology Information (KISTI), National Science & Technology Information Service (NTIS)

向作者/读者索取更多资源

The study assessed the performance of a newly developed AVE-562 analyzer for the vision-based detection of eggs of Clonorchis sinensis in stool samples in Korea, with a correct identification rate of 66.6% and a sensitivity of 36.4%. Although the AVE-562 analyzer enables rapid and convenient stool examination, its sensitivity needs to be improved.
Stool examination is the gold standard for the detection of intestinal parasites. We assessed the performance of a newly developed AVE-562 analyzer (AVE Science & Technology Co., Hunan, China) for the vision-based detection of eggs of Clonorchis sinensis-the most common intestinal parasite in Korea-in stool samples. In total, 30 stool samples with a high or low egg count or without eggs (as negative control samples) (N=10 each) were prepared and analyzed. The performance of the AVE-562 analyzer was compared with that of the formalin-ether concentration (FEC) method. The overall correct identification rate of the AVE-562 analyzer based on FEC results was 66.6%. The sensitivity, specificity, positive predictive value, and negative predictive value of the AVE-562 analyzer for detecting C. sinensis eggs were 36.4%, 100.0%, 100.0%, and 73.1%, respectively. The average time required to run five tests simultaneously was 27 min using the AVE-562 analyzer and 58 min using the FEC method. Although the AVE-562 analyzer enables rapid and convenient stool examination, its sensitivity needs to be improved, particularly considering the prevalence of low-burden C. sinensis infection in Korea.