4.6 Article

Elimination of the non-spectral matrix interferences in the analysis of bio-liquids by mass spectrometry with inductively coupled plasma

期刊

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.sab.2020.106034

关键词

ICP-MS; Non-spectral interferences; Standard and robust instrumental parameters; Bio-liquids

资金

  1. Russian Foundation for Fundamental Research (RFFR) [20-03-00354-a]

向作者/读者索取更多资源

The impact of nitric acid, inorganic salts, and organic compounds on the analytical signals in mass spectrometry analysis of whole blood, serum, and urine has been studied. Non-spectral interferences were found to affect analytes with ionization potentials above 9.0 eV the most, and strategies for their elimination through tuning of the instrumental parameters and selection of internal standards were established. Two approaches were identified for eliminating non-spectral interferences, each with its own advantages in terms of sensitivity and flexibility. These strategies were demonstrated to be efficient in the analysis of standard reference materials and bio-liquids of patients suffering from different forms of SIRS. Concentrations of Cu and Zn were highlighted as diagnostically important features.
Systematic investigations of the impact of nitric acid, inorganic salts and organic compounds on the decrease in the analytical signals in the analysis of whole blood, serum and urine by mass spectrometry with inductively coupled plasma are performed. Two strategies for the sample pretreatment - dilution by different organic mixtures and acidic mineralization - are investigated. The maximum effect of the nitric acid and salts on the results of the determination of the diagnostically important elements P, Mn, Co, Cu, Zn, As, Se, Cd and Pb were found using synthetic model solutions. Analytes with ionization potentials (IP) above 9.0 eV are found to be affected most by these non-spectral interferences. The possibility of avoiding such interferences through simply specific tuning of the instrumental parameters is investigated. It is found that the greatest reduction of the interferences is seen for an increase in the RF power and a decrease in the velocity of the nebulized argon flow. Various strategies for selection of an internal standard (IS) for different tunings of the instrument are found. Two approaches for elimination of the non-spectral interferences are established: i) combination of a standard set of instrumental parameters and the use of the IS with IPs matched to the IPs of the analytes; ii) combination of robust instrumental parameters and the use of a single IS with any mass and IP. The first approach provides higher sensitivity but requires the use of different IS for different analytes. The second approach provides a much more flexible analytical performance for the routine determination of the various sets of analytes but at the expense of a modest decrease in the analytical sensitivity. The efficiency of the developed strategy is demonstrated by the analysis of standard reference materials of whole blood and urine and applied to the analysis of bio-liquids of patients suffering from different forms of a systematic inflammatory response syndrome (SIRS). The concentrations of Cu and Zn were found to be diagnostically important features.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据