Room temperature laser power standard using a microfabricated, electrical substitution bolometer

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 92, 期 2, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/5.0032366

关键词

-

资金

  1. AFMETCAL calibration coordination group (CCG) [CCG-674]
  2. NIST
  3. Jenny and Antti Wihuri Foundation, Finland

向作者/读者索取更多资源

A room temperature electrical substitution radiometer is described for measuring laser power as an absolute standard, achieving high accuracy over a wide range of wavelengths, beam diameters, and powers. The instrument uses a silicon microfabricated bolometer with carbon nanotube absorbers and commercial surface mount thermistors. Validation was done against NIST and AFMETCAL standards.
The design and performance of a room temperature electrical substitution radiometer for use as an absolute standard for measuring continuous-wave laser power over a wide range of wavelengths, beam diameters, and powers are described. The standard achieves an accuracy of 0.46% (k = 2) for powers from 10 mW to 100 mW and 0.83% (k = 2) for powers from 1 mW to 10 mW and can accommodate laser beam diameters (1/e(2)) up to 11 mm and wavelengths from 300 nm to 2 mu m. At low power levels, the uncertainty is dominated by sensitivity to fluctuations in the thermal environment. The core of the instrument is a planar, silicon microfabricated bolometer with vertically aligned carbon nanotube absorbers, commercial surface mount thermistors, and an integrated heater. Where possible, commercial electronics and components were used. The performance was validated by comparing it to a National Institute of Standards and Technology primary standard through a transfer standard silicon trap detector and by comparing it to the legacy C-series standards in operation at the U.S. Air Force Metrology and Calibration Division (AFMETCAL).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据