期刊
RADIATION MEASUREMENTS
卷 141, 期 -, 页码 -出版社
PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.radmeas.2021.106519
关键词
Individual monitoring service; Natural background radiation; Thermoluminescent detector
The average daily natural background dose was calculated using a statistical approach from personal dosemeter records of professionally exposed workers in Switzerland, with values varying depending on geographic location. Measurements with control dosemeters at customer sites were consistent with computed values, but geographic variation and inaccurate correction may result in dose record discrepancies. Up to 3% of dose records during quarterly surveillance periods are expected to be false positives or negatives.
The average daily natural background dose was computed with a statistical approach from the dose records of personal dosemeters worn by professionally exposed workers in Switzerland. The values ranged between 1.3-3.0 mu Sv/day depending on the geographic location of the customers site. The highest values were found in the Lepontine Alps, the lowest in the Swiss Plateau. The results from measurements with control dosemeters at 31 customer sites were in agreement with the computed values. The geographic variation in the natural background and an inaccurate correction are likely to give rise to differences in dose records by up to 0.1 mSv/month. For quarterly surveillance periods up to 3% of the dose records are expected to be false positives or negatives.