4.6 Article

Intraoperative laparoscopic detection of sentinel lymph nodes with indocyanine green and superparamagnetic iron oxide in a swine gallbladder cancer model

期刊

PLOS ONE
卷 16, 期 3, 页码 -

出版社

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0248531

关键词

-

资金

  1. Project for Medical Device Development from the Japan Agency for Medical Research and Development (AMED) [JP18he0902010h0004]

向作者/读者索取更多资源

In this study, a dual tracer laparoscopic technique using ICG and SPIO was found to be feasible for detecting SLNs of gallbladder cancer in a swine model. The results showed that SPIO nanoparticles more accurately identified each SLN, similar to the standard method using radioisotopes.
Mapping of sentinel lymph nodes (SLNs) can enable less invasive surgery. However, mapping is challenging for cancers of difficult-to-access visceral organs, such as the gallbladder, because the standard method using radioisotopes (RIs) requires preoperative tracer injection. Indocyanine green (ICG) and superparamagnetic iron oxide (SPIO) have also been used as alternative tracers. In this study, we modified a previously reported magnetic probe for laparoscopic use and evaluated the feasibility of detecting SLNs of the gallbladder using a laparoscopic dual tracer method by injecting ICG and SPIO into five swine and one cancer-bearing swine. The laparoscopic probe identified SPIO nanoparticles in the nodes of 4/5 swine in situ, the magnetic field counts were 2.5-15.9 mu T, and fluorescence was detected in SLNs in all five swine. ICG showed a visual lymph flow map, and SPIO more accurately identified each SLN with a measurable magnetic field quite similar to the RI. We then developed an advanced gallbladder cancer model with lymph node metastasis using recombination activating gene 2-knockout swine. We identified an SLN in the laparoscopic investigation, and the magnetic field count was 3.5 mu T. The SLN was histologically determined to be one of the two metastatic lymph nodes. In conclusion, detecting the SLNs of gallbladder cancer in situ using a dual tracer laparoscopic technique with ICG and SPIO was feasible in a swine model.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据