Characterization of passive dosimeters in proton pencil beam scanning ? A EURADOS intercomparison for mailed dosimetry audits in proton therapy centres

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2021.01.073

关键词

Alanine; Luminescence dosimetry; Mail auditing; Proton therapy

向作者/读者索取更多资源

The lack of mailed dosimetry audits of proton therapy centres in Europe has led researchers to compare existing passive detector systems in therapeutic pencil beam scanning. The study found that alanine detectors have the lowest energy dependence for different parameters of therapeutic pencil beam scanning and are suitable for mail auditing in proton therapy. The luminescence detector systems show promise, but more careful calibration and corrections are needed for implementation as part of a mailed dosimetry audit system.
The lack of mailed dosimetry audits of proton therapy centres in Europe has encouraged researchers of EURADOS Working Group 9 (WG9) to compare response of several existing passive detector systems in therapeutic pencil beam scanning. Alanine Electron Paramagnetic Resonance dosimetry systems from 3 different institutes (ISS, Italy; UH, Belgium and IFJ PAN, Poland), natLiF:Mg, Ti (MTS-N) and natLiF:Mg, Cu, P (MCP-N) thermoluminescent dosimeters (TLDs), GD-352M radiophotoluminescent glass dosimeters (RPLGDs) and Al2O3:C optically stimulated dosimeters (OSLDs) were evaluate. Dosimeter repeatability, batch reproducibility and response in therapeutic Pencil Beam Scanning were verified for implementation as mail auditing system. Alanine detectors demonstrated the lowest linear energy transfer (LET) dependence with an agreement between measured and treatment planning system (TPS) dose below 1%. The OSLDs measured on average a 6.3% lower dose compared to TPS calculation, with no significant difference between varying modulations and ranges. Both GD-352M and MCP-N measured a lower dose than the TPS and luminescent response was dependent on the LET of the therapeutic proton beam. Thermoluminescent response of MTS-N was also found to be dependent on the LET and a higher dose than TPS was measured with the most pronounced increase of 11%. As alanine detectors are characterized by the lowest energy dependence for different parameters of therapeutic pencil beam scanning they are suitable candidates for mail auditing in proton therapy. The response of luminescence detector systems have shown promises even though more careful calibration and corrections are needed for its implementation as part of a mailed dosimetry audit system.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据