Rapid diagnostic testing for SARS-CoV-2: Validation and comparison of three point-of-care antibody tests

期刊

JOURNAL OF MEDICAL VIROLOGY
卷 93, 期 7, 页码 4592-4596

出版社

WILEY
DOI: 10.1002/jmv.26913

关键词

COVID-19; point-of-care test; SARS-CoV-2; sensitivity; specificity; validation

类别

向作者/读者索取更多资源

The study evaluated three point-of-care antibody tests for detecting IgG and IgM against SARS-CoV-2, with similar performance in sensitivity for IgG but lower sensitivity for IgM. The tests showed potential for accurate use in routine clinical practice.
With the emergence of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), a need for diagnostic tests has surfaced. Point-of-care (POC) antibody tests can detect immunoglobulin (Ig) G and M against SARS-CoV-2 in serum, plasma, or whole blood and give results within 15 min. Validation of the performance of such tests is needed if they are to be used in clinical practice. In this study, we evaluated three POC antibody tests. Convalescent serum samples from 47 reverse transcription-polymerase chain reaction (RT-PCR) verified patients with coronavirus disease 2019 (COVID-19) collected at least 28 days post RT-PCR diagnosis as well as 50 negative pre-COVID-19 controls were tested. The three tests (denoted the J-, N-, and Z-tests) displayed the sensitivities of 87%, 96%, and 85%, respectively, for the detection of IgG. All tests had the same specificity for IgG (98%). The tests did not differ significantly for the detection of IgG. The sensitivities for IgM were lower (15%, 67%, and 70%) and the specificities were 90%, 98%, and 90%, respectively. The positive and negative predictive values were similar among the tests. Our results indicate that these POC antibody tests might be accurate enough to use in routine clinical practice.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据