Different diagnostic tests are being developed for detecting SARS-CoV-2, but discrepancies in results can occur, causing confusion. The study found that FilmArray RP v2.1 can detect SARS-CoV-2 with accuracy comparable to RT-qPCR, especially in early and intermediate phases of infection.
Various diagnostic tests utilizing different principles are currently under development for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). However, these tests can occasionally produce discrepant results, causing confusion in their interpretation. Here, we evaluated the performance and features of three diagnostic assays: quantitative reverse transcription polymerase chain reaction (RT-qPCR), FilmArray Respiratory Panel (RP) v2.1, and the LUMIPULSE antigen test. Twenty-seven serial nasopharyngeal swabs were collected from a prolonged viral shedding patient who had been hospitalized for 51 days. We examined the SARS-CoV-2 detection rates of the three tests. The overall agreement rate was 81% between RT-qPCR and FilmArray RP v2.1, 63% between the antigen test and FilmArray RP v2.1, and 59% between the antigen test and RT-qPCR. We obtained concordant results in samples with high viral loads (low threshold cycle values) by all three tests. RT-qPCR and FilmArray RP v2.1 accurately detected SARS-CoV-2 at the early to intermediate phases of infection, but the results varied at the late phase. The antigen test also produced a positive result at the early phase but varied at the intermediate phase and consistently produced negative results at late phase of infection. These results demonstrated FilmArray RP v2.1 could detect SARS-CoV-2 with accuracy comparable to RT-qPCR. Further, there were discrepant results using different types of diagnostic tests during the clinical course of prolonged viral shedding patient. We provided insights into how to utilize different types of kits to assess and manage SARS-CoV-2 infections. (C) 2020 Japanese Society of Chemotherapy and The Japanese Association for Infectious Diseases. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据