Out-of-field doses from radiotherapy using photon beams: A comparative study for a pediatric renal treatment

期刊

出版社

WILEY
DOI: 10.1002/acm2.13182

关键词

advanced radiotherapy; peripheral doses; radiochromic films; treatment planning system

向作者/读者索取更多资源

This study demonstrates that advanced radiotherapy techniques deliver higher out-of-field doses compared to conventional techniques, with differences increasing with distance to target. Tomotherapy delivers lower doses than VMAT due to additional shielding, and the Halcyon system has lower peripheral doses. Tomotherapy is suitable for out-of-field dose determination up to 30 cm, while Eclipse TPS underestimates those doses.
Purpose First, this experimental study aims at comparing out-of-field doses delivered by three radiotherapy techniques (3DCRT, VMAT (two different accelerators), and tomotherapy) for a pediatric renal treatment. Secondly, the accuracy of treatment planning systems (TPS) for out-of-field calculation is evaluated. Methods EBT3 films were positioned in pediatric phantoms (5 and 10 yr old). They were irradiated according to four plans: 3DCRT (Clinac 2100CS, Varian), VMAT (Clinac 2100CS and Halcyon, Varian), and tomotherapy for a same target volume. 3D dose determination was performed with an in-house Matlab tool using linear interpolation of film measurements. 1D and 3D comparisons were made between techniques. Finally, measurements were compared to the Eclipse (Varian) and Tomotherapy (Accuray) TPS calculations. Results Advanced radiotherapy techniques (VMATs and tomotherapy) deliver higher out-of-field doses compared to 3DCRT due to increased beam-on time triggered by intensity modulation. Differences increase with distance to target and reach a factor of 3 between VMAT and 3DCRT. Besides, tomotherapy delivers lower doses than VMAT: although tomotherapy beam-on time is higher than in VMAT, the additional shielding of the Hi-Art system reduces out-of-field doses. The latest generation Halcyon system proves to deliver lower peripheral doses than conventional accelerators. Regarding TPS calculation, tomotherapy proves to be suitable for out-of-field dose determination up to 30 cm from field edge whereas Eclipse (AAA and AXB) largely underestimates those doses. Conclusion This study shows that the high dose conformation allowed by advanced radiotherapy is done at the cost of higher peripheral doses. In the context of treatment-related risk estimation, the consequence of this increase might be significative. Modern systems require adapted head shielding and a particular attention has to be taken regarding on-board imaging dose. Finally, TPS advanced dose calculation algorithms do not certify dose accuracy beyond field edges, and thus, those doses are not suitable for risk assessment.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据