4.4 Article

Sample Length Dependence of Magnetic Moment Detected by a SQUID Magnetometer Using Nickel Cylinders

期刊

IEEE TRANSACTIONS ON MAGNETICS
卷 57, 期 2, 页码 -

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TMAG.2020.3006209

关键词

Calibration; nickel; sample size; superconducting quantum interference device (SQUID); yttrium iron garnet (YIG)

资金

  1. Nanotechnology Platform Program (Molecule and Material Synthesis) of the Ministry of Education, Culture, Sports, Science and Technology (MEXT), Japan
  2. JSPS KAKENHI [JP17K05918]

向作者/读者索取更多资源

The study reported measurements in a second-order gradiometer superconducting quantum interference device (SQUID) magnetometer of pure nickel (Ni) cylinders. It found that the magnetic moment per unit mass of the Ni cylinder decreases with sample length, with the extent of decrease depending on the type of magnetometer used and the length of the second-order gradiometer superconducting coil.
We report here on sample length dependence measurements in a second-order gradiometer superconducting quantum interference device (SQUID) magnetometer of pure nickel (Ni) cylinders, having a diameter of 1 mm and a length of 1-7 mm. The measured magnetic moment per unit mass significantly decreases with the sample length of the Ni cylinder, when a magnetic field of 7 T was applied parallel to the long axis of the cylinder. As an example, the magnetic moment per unit mass for the 5 mm Ni cylinder is between 97% and 94% of that for the 1 mm Ni cylinder, the number depending on the specific type of magnetometer used. The magnitude of decrease also depends on the length of the second-order gradiometer superconducting coil which differs with different models of SQUID magnetometers.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据