EUROPEAN RADIOLOGY
卷 31, 期 9, 页码 6793-6801出版社
SPRINGER
DOI: 10.1007/s00330-021-07824-x
关键词
Computed tomography; X-ray; Pelvic bones; Tin; Radiation dosage
资金
- Universitat Zurich
Ultra-low-dose CT with tin filtration for the osseous pelvis showed a significant dose reduction of 84% compared to standard CT, while maintaining similar imaging quality for assessing bone anatomy and pathologies. Additionally, ULD-CT demonstrated a markedly superior dose efficiency compared to standard CT.
Objectives To compare ultra-low-dose CT (ULD-CT) of the osseous pelvis with tin filtration to standard clinical CT (CT), and to assess the quality of computed virtual pelvic radiographs (VRs). Methods CT protocols were optimized in a phantom and three pelvic cadavers. Thirty prospectively included patients received both standard CT (automated tube voltage selection and current modulation) and tin-filtered ULD-CT of the pelvis (Sn140kV/50mAs). VRs of ULD-CT data were computed using an adapted cone beam-based projection algorithm and were compared to digital radiographs (DRs) of the pelvis. CT and DR dose parameters and quantitative and qualitative measures (1 = worst, 4 = best) were compared. CT and ULD-CT were assessed for osseous pathologies. Results Dose reduction of ULD-CT was 84% compared to CT, with a median effective dose of 0.38 mSv (quartile 1-3: 0.37-0.4 mSv) versus 2.31 mSv (1.82-3.58 mSv; p < .001), respectively. Mean dose of DR was 0.37 mSv (+/- 0.14 mSv). The median signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of bone were significantly higher for CT (64.3 and 21.5, respectively) compared to ULD-CT (50.4 and 18.8; p <= .01), while ULD-CT was significantly more dose efficient (figure of merit (FOM) 927.6) than CT (FOM 167.6; p < .001). Both CT and ULD-CT were of good image quality with excellent depiction of anatomy, with a median score of 4 (4-4) for both methods (p = .1). Agreement was perfect between both methods regarding the prevalence of assessed osseous pathologies (p > .99). VRs were successfully calculated and were equivalent to DRs. Conclusion Tin-filtered ULD-CT of the pelvis at a dose equivalent to standard radiographs is adequate for assessing bone anatomy and osseous pathologies and had a markedly superior dose efficiency than standard CT.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据