4.7 Article

Limiting radiation exposure during prostatic arteries embolization: influence of patient characteristics, anatomical conditions, and technical factors

期刊

EUROPEAN RADIOLOGY
卷 31, 期 9, 页码 6471-6479

出版社

SPRINGER
DOI: 10.1007/s00330-021-07844-7

关键词

Benign prostatic hyperplasia; Cone beam computed tomography; Interventional radiology; Prostatic artery embolization; Radiation exposure

向作者/读者索取更多资源

The study found that using large display monitors and cone beam computed tomography can significantly reduce radiation exposure during prostatic arteries embolization (PAE) for benign prostatic hyperplasia. Lower patient BMI, shorter fluoroscopy time, CBCT use, and decreased number of acquisitions were associated with reduced air kerma dosage.
Objective To assess the influence of patient characteristics, anatomical conditions, and technical factors on radiation exposure during prostatic arteries embolization (PAE) performed for benign prostatic hyperplasia. Materials and methods Patient characteristics (age, body mass index (BMI)), anatomical conditions (number of prostatic arteries, anastomosis), and technical factors (use of cone beam computed tomography (CBCT), large display monitor (LDM), and magnification) were recorded as well as total air kerma (AK), dose area product (DAP), fluoroscopy time (FT), and number of acquisitions (NAcq). Associations between potential dose-influencing factors and AK using univariate analysis and a multiple linear regression model were assessed. Results Forty-one consecutive men (68 +/- 8 years, min-max: 40-76) were included. LDM and CBCT decreased the use of small field of view with 13.9 and 3.8% respectively, both p < 0.001. The use of a LDM significantly reduced AK (1006.6 +/- 471.7 vs. 1412 +/- 754.6 mGy, p = 0.02), DAP (119.4 +/- 64.4 vs. 167.9 +/- 99.2, p = 0.04), FT (40.4 +/- 11.5 vs. 53.6 +/- 25.5 min, p = 0.01), and NAcq (16.3 +/- 6.3 vs. 18.2 +/- 7, p = 0.04). In multivariate analysis, AK reduction was associated with lower patient BMI (beta = 0.359, p = 0.002), shorter FT (beta = 0.664, p < 0.001) and CBCT use (beta = - 0.223, p = 0.03), and decreased NAcq (beta = 0.229, p = 0.04). Conclusion LDM and CBCT are important technical dose-related factors to help reduce radiation exposure during PAE, and should be considered in standard practice.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据